Thanks. This is very informative.
Basically it confirms the 10uA limit for any electrode.
Further it specifies the voltage at which the insulation (e.g. optocouplers) would be tested.
"Insulation breakdown
IEC601-1 Clause 20
This test is actually a series of tests designed to test the strength of insulation. Basically you apply a high voltage, as much as
5500 volts, to an insulation barrier to determine if it will "withstand" the voltage and not short across.
..."