Hi,
I have changed my comEEG prototype back, as it has been before
and done some noise measurements.
The picture shows the 50uVp-p calibration signal to the right
(height in pixel: 167) and the noise of the shorted input to the
left side. The noise is 4 pixel high: That means about 1.2uVp-p
The 555 PWM-ADC contributes half of the noise, the rest is from the
amplifier stage.
Regards,
Joerg